For a better experience please change your browser to CHROME, FIREFOX, OPERA or Internet Explorer.
Please contact Us / WhatsApp
+91 8218684097

INDUCTIVE CONDUCTIVITY TRANSMITTER – FLS C6.30

FLS C6.30 is a family of inductive conductivity transmitters which are made by a 4-20mA output device (two wire technology) integrated onto an inductive conductivity sensor. This type of measurement technology allows a broad range of applications especially to measure high values of conductivity (till 1000 mill siemens) in aggressive fluids (C-PVC is the only wetted material). As no electrodes are directly in contact with liquid, a reliable and stable measurement is guaranteed for a long operating time. A proper automatic temperature compensation (ATC) is granted by a Pt100 integrated into instrument body. The isolated 4-20mA output is perfect for direct connections to PLCs or data loggers without any extra interfaces. Transmitter as well as temperature sensor are supplied already calibrated.

APPLICATIONS

• Water treatment
• Waste water treatment
• Cooling towers
• Scrubber systems
• Metal finishing Coating and corrosion fluid measurement

MAIN FEATURES

• Corrosion & coating resistant
• Compact transmitter
• No calibration required
• Easy installation
• Pt100 sensor integrated
• Suitable for submersible installation

Technical Data

General

• Body Material: C-PVC
• Body length: 207mm
• Enclosure: IP68
• Power supply: 10-30 VDC regulated
• Max Power Consumption: <22mA
• Max Working pressure/Working
Temperature:
– 10bar (145psi) @ 25°C (77°F)
– 6bar (87psi) @ 50°C (122°F)
• Process connection: 1 ½” male NPT

Standards & Approvals

• Manufactured under ISO 9001
• Manufactured under ISO 14001
• CE
• RoHS Compliant
• EAC

DIMENSIONS

ORDERING DATA

C6.30 Inductive Conductivity Transmitters

Part No. Description /Name Applications/
Operative Range
Cell Constant Connection Installation Weight (gr.)
C6.30.01 C-PVC Inductive
Conductivity Transmitter
with Temperature Sensor
included
0-10 mS/cm 3 mt 1 1/2″ NPT
male threaded
550
C6.30.02 C-PVC Inductive
Conductivity Transmitter
with Temperature Sensor
include
0-100 mS/cm 3 mt 1 1/2″ NPT
male threaded
550
C6.30.03 C-PVC Inductive
Conductivity Transmitter
with Temperature Sensor
include
0-1000 mS/cm 3 mt 1 1/2″ NPT
male threaded
550

Flow Measurement System

GET THE LATEST FLOW MEASUREMENT SYSTEM PRICES ON YOUR EMAIL

Request Price
Top