STAINLESS STEEL CONDUCTIVITY SENSOR – FLS C100-300

The FLS conductivity sensors with stainless steel electrodes (C100 series) are designed for agriculture application and for light industrial application obviously where sample conditions allow a steel using (water treatment, foods industry and others). This type of sensors are characterized by a considerable ratio performance/price. The combination of temperature sensor with the ATC (Automatic Temperature Compensation) function of the monitor/transmitter allows to get a precise measurement. Moreover a wide number of cell constants grants to choose the best item for specificapplication.C300 serie has been designed for ultrapure water monitoring (certified cell constant 0,01)and for waste water application (cell constant 10). C300 sensors are completely made in SS ensuring suitability for a wide range of applications.

APPLICATIONS

• Agriculture and fertilizing system
• Water treatment
• Foods industry
• Aquaculture
• Ultrapure water application: production and use

MAIN FEATURES

• Stainless steel measuring surfaces
• Considerable ratio performance/price
• Temperature sensor included
• Wide range of cell constant
• Rugged sensor body in PP (C100)
• Sensor in SS completely (C300)
• C300.001TCCK with certified cell constant

Technical Data

– measuring surface: stainless steel 316

Standards & Approvals

• Manufactured under ISO 9001
• Manufactured under ISO 14001
• CE
• RoHS Compliant
• EAC

Optimal Ranges

cell constant0,010,10,2110
conductivity range0,055 ÷ 20 µS/cm0,5 ÷ 200 µS/cm1 ÷ 400 µS/cm0,005 ÷ 10 mS/cm0,5 ÷ 200 mS/cm
resistivity range18,18 ÷ 0,05 MΩ*cm2000 ÷ 5 kΩ*cm1 ÷ 0,0025 MΩ*cm200÷ 0,1 kΩ*cm2 ÷ 0,005 kΩ*cm

DIMENSIONS

A C100.01 TC
B C100.1 TC, C100.02 TC
C C300.001 TC
D C300.10 TC

2 PP body ½“ male threaded BSP
3 Stainless Steel electrodes
4 4-Poles connector

5 PP body ¾” male threaded BSP
6 Cable: 3m (10 ft.)

ORDERING DATA

C100 Stainless Steel Electrodes Conductivity Sensors

Part No.Description /NameApplications/
Operative Range
Cell ConstantConnectionInstallationWeight
(gr.)
C100.01TCPP body sensor with SS
Conductivity Electrode
and Temperature Sensor
included
0.1 µS/cm to 2000 µS/cm
(10 MΩ*cm
to 500 Ω *cm)
0,14 poles
connector
¾” male
BSP (parallel threaded)
350
C100.02TCPP body sensor with SS
Conductivity Electrode
and Temperature Sensor
included
0.2 µS/cm
to 4000 µS/cm
0.24 poles
connector
¾” male
BSP (parallel threaded)
350
C100.1TCPP body sensor with SS
Conductivity Electrode
and Temperature Sensor
included
1 µS/cm
to 20000 µS/cm
14 poles
connector
¾” male
BSP (parallel threaded)
350

C300 Stainless Steel Electrodes Conductivity Sensors

Part No.Description /NameApplications/
Operative Range
Cell ConstantConnectionInstallationWeight
(gr.)
C300.001TCBody sensor and conductivity electrodes in SS0,055 µS/cm to 200 µS/cm (18,2MΩ*cmto 5KΩ*cm)0,013 mPP body ½” male (EG12SS option)150
C300.001TCCKBody sensorand conductivity electrodesin SS with certifiedcell constant0,055 µS/cm to 200 µS/cm (18,2MΩ*cmto 5KΩ*cm)0,013 mPP body ½” male (EG12SS option)150
C300.10TCBody sensor and conductivity electrodes in SS10 µS/cm to 200000 µS/cm103 mPP body ½” male (EG12SS option)150